How FDA classifies medical devices

A short primer for students, career changers, and coaching clients who are new to medtech. Class follows risk, and most of what comes next follows class.

FDA sorts medical devices into three classes by risk. Class I carries the lowest risk and Class III the greatest, and each class comes with its own level of regulatory control. Learn this first.

What counts as a medical device

Section 201(h)(1) of the Federal Food, Drug, and Cosmetic Act covers an instrument, apparatus, implement, machine, implant, in vitro reagent, or similar article, including a component, part, or accessory, that is intended for use in the diagnosis, cure, mitigation, treatment, or prevention of disease, or intended to affect the structure or any function of the body. Unlike a drug, a device does not achieve its primary intended purposes through chemical action within or on the body and does not depend on being metabolized.

Who regulates devices

The Center for Devices and Radiological Health (CDRH) is the FDA center that oversees medical devices and radiation-emitting products. FDA has established classifications for approximately 1,700 different generic types of devices and grouped them into 16 medical specialty panels. Classification depends on the intended use of the device and on its indications for use.

Diagram of FDA medical device classes

The diagram is a simplified 2020 picture, and its timelines are rough ranges.

The three classes and their controls

  • Class I (low to moderate risk) is subject to general controls, such as establishment registration, device listing, and adverse event reporting.
  • Class II (moderate to high risk) adds special controls, which are usually device-specific and can include performance standards and postmarket surveillance.
  • Class III (high risk) is subject to general controls and premarket approval. Devices that support or sustain human life are the familiar example.

Manufacturing quality has its own rule. The Quality Management System Regulation (QMSR) is the amended 21 CFR Part 820, which now incorporates the international standard ISO 13485:2016. It took effect on February 2, 2026.

The main premarket pathways

Class points to pathway. Most Class I and some Class II devices are exempt from premarket notification. A Class I or II device that is not exempt needs a 510(k), and a Class III device generally needs premarket approval.

  • A 510(k), or premarket notification, demonstrates that the new device is substantially equivalent to a predicate device.
  • A De Novo request asks FDA to classify a novel device that has no legally marketed predicate into Class I or II.
  • Premarket approval (PMA) requires valid scientific evidence demonstrating reasonable assurance of safety and effectiveness.
  • A Humanitarian Device Exemption (HDE) is for a device intended to benefit patients with a disease or condition that affects not more than 8,000 people in the United States per year. It is exempt from the effectiveness requirements of sections 514 and 515 of the Act.

Where software and AI fit

Software can be a device on its own. The International Medical Device Regulators Forum defines software as a medical device (SaMD) as software intended for one or more medical purposes that performs them without being part of a hardware medical device. Section 520(o) of the Act excludes certain software functions from the device definition.

AI-enabled devices use the same pathways, and FDA publishes a list of those authorized for marketing in the United States. It has also issued final guidance on predetermined change control plans, which describe planned modifications to AI-enabled device software in the marketing submission.

Where cybersecurity enters premarket review

Cybersecurity is written into the statute now. Section 524B of the Act took effect on March 29, 2023. It applies to a cyber device, meaning a device that includes software, can connect to the internet, and has technological characteristics that could be vulnerable to cybersecurity threats. The sponsor of a premarket submission must submit a plan to monitor, identify, and address postmarket vulnerabilities, maintain processes that provide reasonable assurance the device is cybersecure, and provide a software bill of materials.

How to use this page

This page is educational material. Decisions about a specific device belong with the company’s own regulatory team. If you are preparing for a medtech interview, look up the company’s product in FDA’s classification database before you walk in.